SA2100 Portable Pulse Oximeter
Specifications
Performance
Scope of measurement
SpO2: 35% - 100%
Pulse rate: 30-250 bpm (beats per minute)
Precision
SpO2: 70% - 100%: +_2 digits (%)
35% - 69%: unspecified
Pulse rate: 30- 250 +_3 digits
Data Update period: less than 2 seconds
Averaging: 4 for SpO2; 8 for Pulse rate
Alarm condition delay: Less than 8 seconds for spo2;
Less than 16 seconds for pulse rate
Alarm signal generation delay: Less than 1 seconds for spo2 and pulse rate
Probe type
Probe: Rossmax PA100, Solaris S100A
Extension cord: Rossmax PE100, Solaris AC031-113
Electrical Specifications
Battery (4”AA” Alkaline batteries)
Battery capability: Can be used for 15 hours continuously depending on the type
of battery used
Environmental conditions
Operating temperature 5°C - 40°C (41°F - 104°F)
Storage temperature
Packaged: -20°C - 70°C (-4°F - 158°F)
Naked: -20°C - 60°C (-4°F - 140°F)
Relative humidity: 15% - 95% (no condensing)
Physical characteristics
Weight: 0.15Kg (excluding battery and probe) (0.33 ponds)
Size: 5.70 in. x 2.85 in. x 0.89 in. (14.5cm x 7.25cm x 2.25cm)
|